Dissolved oxygen sensor film using ruthenium polypyridine complex as luminophore

ผู้แต่ง

  • Chanissnun Su-Ngam Faculty of Science, Kasetsart University and Center for Advanced Studies in Tropical Natural Resources National Research UniversityKasetsart University (CASTNAR,NRU-KU)
  • San H. Thang CSIRO Molecular and Health Technologies, Bayview Avenue
  • Atchana Wongchaisuwat Faculty of Science, Kasetsart University and Center for Advanced Studies in Tropical Natural Resources National Research UniversityKasetsart University (CASTNAR,NRU-KU)
  • Ladda Meesuk Faculty of Science, Kasetsart University and Center for Advanced Studies in Tropical Natural Resources National Research UniversityKasetsart University (CASTNAR,NRU-KU)

คำสำคัญ:

Ruthenium polypyridine complex, RAFT, Dissolved oxygen, Luminophore

บทคัดย่อ

Ruthenium polypyridine complex was synthesized by using RAFT process. The Ruthenium was immobilized in 2 silicone matrices, polydimethylsiloxane(PDMS) and KE 200 (silicone elastomer base)/CX200 (curing agent). KE 200/CX200 is the alcohols curing type silicone. Both sensor films showed absorption peak at 460 nm and  luminescence peak at 610 nm. The KE200/CX200 film gave linear response to dissolved oxygen up to 80% O2 (18 ppm) follow Stern-Volmer equation with R2 of 0.990. The measured luminescence intensity decreased with increasing dissolved oxygen water temperature from 25 to 70°c. After being irradiated with 460 nm light source for 3 hrs, the sensor film still showed good stability.

Downloads

Download data is not yet available.

เอกสารอ้างอิง

Di Marco, G. and Lanza, M. (2000). Optical solidstate oxygen sensors using metalloporphyrin complexes immobilized in suitable polymeric matrices. Sens. Actuators,B.63:42-48.

Chu, C.S. and Lo, Y.L. (2008). Ratiometric fiber-optic oxygen sensors based on sol-gel matrix doped with metalloporphyrin and 7- amino-4-trifluoromethyl coumarin. Sens. Actuators, B. 134(2) : 711-717.

Jones, R. G. (1995). Silicon-Containing Polymers. Cambridge : RoyalSociety of Chemistry.

Choi, M.M.F. and Xiao, D. (2000). Single standard calibration for an optical oxygen sensor based on luminescence quenching of a ruthenium complex. Anal. Chim. Acta. 403 : 57–65.

Chen, M., Ghiggino, K.P., Launikonis, A., Mau, A.W.H., Rizzardo, E., Sasse, W.H.F., Thang, S.H. and Wilson, G.J. (2003). RAFT synthesis of linear and star-shaped light harvesting polymers using di- and hexafunctional ruthenium polypyridine reagents. J. Mater. Chem. 13 : 2696-2700.

McDonald, J.C., Duffy, D.C., Anderson, J.R., Chiu, D.T., Wu, H. and Whitesides, G. M. (2000). Fabrication of microfluidic systems in poly (dimethylsiloxane). Electrophoresis. 21 : 27-40.

Lakowicz, J.R. (2006). Principle of Fluorescence Spectroscopy. 3rd ed. New York : Springer.